§ 35.2647 - Records of additional technical requirements for mobile remote afterloader units.

Copy with citation
Copy as parenthetical citation

A licensee shall retain a record of each check for mobile remote afterloader units required by § 35.647 for 3 years.

The record must include—

The date of the check;

The manufacturer's name, model number, and serial number of the remote afterloader unit;

Notations accounting for all sources before the licensee departs from a facility;

Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and

The signature of the individual who performed the check.